Magnetic field compensation coil design for magnetoencephalography

Zugehörigkeit
Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany
Kutschka, Hermann;
Zugehörigkeit
Institute of Psychology, Leipzig University, Leipzig, Germany
Doeller, Christian F.;
Zugehörigkeit
Hans-Berger Department of Neurology, University Hospital Jena, Jena, Germany
Haueisen, Jens;
Zugehörigkeit
Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany
Maess, Burkhard

Abstract While optically pumped magnetometers (OPMs) can be attached to the head of a person and allow for highly sensitive recordings of the human magnetoencephalogram (MEG), they are mostly limited to an operational range of approximately 5 nT. Consequently, even inside a magnetically shielded room (MSR), movements in the remnant magnetic field disable the OPMs. Active suppression of the remnant field utilizing compensation coils is therefore essential. We propose 8 compensation coils on 5 sides of a cube with a side length of approximately 2 m which were optimized for operation inside an MSR. Compared to previously built bi-planar compensation coils, the coils proposed in this report are more complex in geometry and achieved smaller errors for simulated compensation fields. The proposed coils will allow for larger head movements or smaller movement artifacts in future MEG experiments compared to existing coils.

Zitieren

Zitierform:
Zitierform konnte nicht geladen werden.

Rechte

Nutzung und Vervielfältigung: